| Corrosive medium: acids, alkalis, chlorides, seawater | | Diaphragms in Hastelloy C, tantalum or gold-plated steel, as Klay offers, or a ceramic sensor, as on the Klay 3000 and CER series — or any instrument behind a diaphragm seal in a resistant material. | — | Any type can be used if every wetted part resists the medium; the materials matter more than the type of instrument. |
|---|
| Hygienic process: food, dairy, beverages, pharmaceuticals | | A flush diaphragm on a hygienic connection leaves no pocket for product or bacteria and can be cleaned in place — the Klay 4000-SAN, for example, carries EHEDG and 3-A. | — | Avoid any instrument with a threaded port or a dead space where product can stay; a gauge or switch needs a hygienic diaphragm seal. |
|---|
| Viscous, crystallising or solids-laden medium: pulp, slurries, sludge, polymers | | A flush diaphragm has no opening to clog — Klay builds flush-diaphragm connections for pulp and paper. | | A standard gauge has a narrow port and a Bourdon tube that block; if you need local indication here, fit the gauge with a diaphragm seal. |
|---|
| Level in a closed or pressurised tank, or pressure drop across a filter | | One instrument measures the difference directly: the pressure at the bottom minus the gas pressure above the liquid, or the pressure before minus after the filter. | | A single transmitter at the bottom of a closed tank reads the liquid plus the gas pressure; subtracting two transmitters adds the errors of both. |
|---|
| Local indication with no power: pumps, compressors, filters, utility lines | | A gauge needs no power and no wiring, and shows the pressure on the spot. | | A transmitter needs loop power, and a display or a control system to read it. |
|---|
| Only an alarm or a trip at a set pressure: high or low pressure protection, interlocks | | The contact switches directly, with no PLC and no programming, and works independently of the control system; versions with SIL 2 certification exist, such as the 12 Series. | — | If you also need the value, use a transmitter with alarms in the control system — or a combined instrument such as the One Series, which is a switch and a transmitter in one. |
|---|
| Very high pressure: hydraulics, high-pressure gas, injection | | Transmitters such as the D22 and D23 reach 1,000 bar and the TX300 30,000 psi; the GS gauge has ranges up to 4,000 bar. | | It measures a difference over a limited span — up to 70 bar on the D35 — not a high gauge pressure. |
|---|
| Very low pressure or vacuum: furnace draft, ventilation, clean rooms | | Small spans in millibar, measured against the atmosphere or another room — the D34 is made for draft and furnace measurements, and the DP-4000 starts at 10 mbar. | | A Bourdon tube is not sensitive enough at a few millibar; that needs a diaphragm or capsule gauge, or a transmitter. |
|---|
| Hazardous areas (ATEX) | | Intrinsically safe transmitters, such as the Klay 2000, 4000 and 8000 with ATEX and IECEx Ex ia, explosion-proof ones such as the TX300, and flameproof switches such as the Delta Mobrey 301. | — | Any type can be used with the right certification and installation; we check it for your zone. |
|---|